Ph · Physics

Can a PM100 Measure Total Power Across 400–800 nm?

Published By the 9to5Science editorial team

At a glance

  1. 1. Check the sensor

    Confirm the attached sensor covers 400–800 nm and can handle the beam’s power and size.

  2. 2. Capture the beam

    Align the beam so the sensor receives all of it, and avoid saturation or stray light.

  3. 3. Account for spectrum

    A single wavelength setting is generally not enough to convert a broadband sensor signal into total power.

  4. 4. Apply a correction

    Use the source spectrum and sensor responsivity, or use a suitably calibrated broadband measurement method.

Advertisement

A Thorlabs PM100 is a meter console used with a compatible optical sensor. It can register a broadband source, but whether its displayed value equals the total power from 400 to 800 nm depends on the sensor’s wavelength response and how the measurement is calibrated.

First, distinguish the quantities: a power sensor normally reports optical power in watts. Irradiance is power per area, in W/m², while radiant intensity is power per solid angle, in W/sr. If by “intensity” you mean the total optical power in the beam, the PM100 reading is the relevant starting point.

Why broadband light needs a correction

Let B=[400,800]B=[400,800] nm be the band of interest, and let PλP_\lambda be the source’s spectral power incident on the sensor (power per unit wavelength). For a photodiode sensor, the measured photocurrent is

I=∫ΛR(λ)Pλ dλ,I=\int_{\Lambda} R(\lambda)P_\lambda\,d\lambda,

where R(λ)R(\lambda) is the sensor responsivity, usually in A/W, and Λ\Lambda covers all wavelengths reaching the sensor that produce a signal. If the source has no power outside BB, or out-of-band light is blocked, this reduces to an integral from 400 to 800 nm.

When you choose a wavelength on the meter, it uses the sensor’s calibration for that wavelength to convert its signal into a power reading. In simplified form, that reading is I/R(λ0)I/R(\lambda_0), where λ0\lambda_0 is the selected wavelength. This is appropriate for light concentrated near λ0\lambda_0. For a broad spectrum, the detector signal is a weighted sum across wavelengths, so dividing by the responsivity at one wavelength may overestimate or underestimate the total.

The PM100 console alone does not determine the source spectrum. Also, the sensor head—not just the console—sets the usable wavelength range and power limits. Check the documentation for the exact PM100 model and sensor you have. Some thermopile sensors have a flatter spectral response than photodiodes over their specified range, but their specifications still govern whether broadband accuracy is adequate.

How to obtain total power

  1. Check the sensor specifications. Verify that its stated wavelength range includes the full interval from 400 to 800 nm. Check its maximum power, aperture, and any wavelength-dependent calibration information.
  2. Measure the beam correctly. Place the sensor where it intercepts the whole beam. Keep the reading within the sensor’s range, minimize stray light, and allow the reading to stabilize as the sensor instructions recommend.
  3. Record the wavelength setting and reading. The chosen wavelength matters for a wavelength-dependent sensor. Do not treat the displayed value as automatically spectrum-integrated just because the source contains many wavelengths.
  4. Correct for the spectrum if needed. Measure or otherwise know the source’s spectral power distribution, and combine it with the sensor’s responsivity data over all wavelengths that contribute to the signal. To obtain only the 400–800 nm power, block out-of-band light or account for it in the calculation. Alternatively, use a broadband sensor whose specified response and calibration meet the accuracy you need.

If the spectrum is known, the correction for the power in band BB can be expressed as

PB=I ∫BPλ dλ∫ΛR(λ)Pλ dλ.P_B=I\,\frac{\int_B P_\lambda\,d\lambda}{\int_{\Lambda} R(\lambda)P_\lambda\,d\lambda}.

Here II is the measured photocurrent, and the denominator accounts for all wavelengths contributing to that current. If only light in BB reaches the sensor, then Λ=B\Lambda=B. In practice, the integrals may be calculated from measured wavelength bins. The meter’s wavelength setting by itself cannot supply this information.

Worked example with a two-band source

Suppose a source has 5 mW at 450 nm and 5 mW at 750 nm, with no other light reaching the sensor. Its actual total power is 10 mW. Assume, for illustration, the sensor responsivity is 0.20 A/W at 450 nm and 0.40 A/W at 750 nm. The photocurrent is

I=(0.20)(0.005)+(0.40)(0.005)=0.003 A=3 mA.I=(0.20)(0.005)+(0.40)(0.005)=0.003\,\mathrm{A}=3\,\mathrm{mA}.

If you set the meter to a wavelength where the sensor responsivity is 0.25 A/W, the simplified displayed value would be

Pdisplay=0.003 A0.25 A/W=0.012 W=12 mW.P_{\mathrm{display}}=\frac{0.003\,\mathrm{A}}{0.25\,\mathrm{A/W}}=0.012\,\mathrm{W}=12\,\mathrm{mW}.

That is 20% higher than the true 10 mW total. The responsivity values here are hypothetical, not specifications for a particular PM100 sensor. The example shows why the source spectrum and the actual sensor’s calibration data matter.

Common mistakes

  • Assuming “visible” means the sensor covers the whole band. The 400–800 nm interval is wider than some sensors’ specified range. A reading outside the calibrated range may be inaccurate even if the sensor still responds.
  • Using the center wavelength as if it were a broadband correction. A single setting is exact only under suitable conditions, such as a narrowband source or a response that is sufficiently flat for the required accuracy.
  • Confusing power with irradiance. A power-meter sensor generally reports the power it intercepts, not power per unit area. To estimate irradiance, divide the measured power by the illuminated area only when the beam and area are defined appropriately.
  • Calling a reading “total visible power” without defining the band. A sensor reading includes whatever wavelengths reach it within its response; it does not automatically isolate 400–800 nm. Light outside that band can contribute too if the sensor responds to it.

For a narrowband source, set the PM100 to the source wavelength and follow the sensor’s calibration procedure. For a broadband source with unknown spectral distribution, use a spectrometer together with calibrated sensor data, or choose a measurement setup specifically specified for broadband total-power measurements.

References

  1. Thorlabs PM100D Optical Power and Energy Meter Console User Guide
  2. Thorlabs Photodiode Power Sensors documentation

Spotted an error? Send a correction. Read how we write and check articles in our editorial policy.